36W Series

Type:ATM036T-P/A

Information

ATM036T-□XXX

XXX = O/P Voltage

= P : C14
A : C8 / C18

Features:

* IEC/EN/ANSI/AAMI ES60601-1 (Edition 3.2)
* EMC : IEC60601-1-2 (Edition 4.1)
* Means of Protection: 2 X MOPP
* 100-240VAC Universal Input
* Regulated Output With Low Ripple Noise 
* Safety Agency Requirements and EMI/EMS Certified
* Modified and Custom Design Available
* 2 Years Warranty

Specifications

Model O/P Voltage O/P Current Watt
ATM036T-□050 5.0V 5.0A 25W
ATM036T-□075 7.5V 4.0A 30W
ATM036T-□090 9.0V 4.0A 36W
ATM036T-□120 12.0V 3.0A 36W
ATM036T-□150 15.0V 2.4A 36W
ATM036T-□180 18.0V 2.0A 36W
ATM036T-□240 24.0V 1.5A 36W

Specifications Details

Input
Voltage 100-240 VAC
Line Frequency 50-60Hz
Current 1.0A-0.45A
Protection Internal Primary Current Fuse
Configuration IEC60320/C8, C14, C18
Output
Load Regulation ± 5% (Typical)
Ripple & Noise ≤2% Vp-p @ Full Load
Transient(Dynamic) Response 0.5mS with 50% Load Change
Start-up Time < 3,000mS
Hold-up Time ≥ 10mS @ Full Load
Rise Time < 50mS
Protection Short Circuit Protection/ Over Voltage Protection/ Over Current Protection
Safety Approvals
Approvals CB / UL / cUL / FCC / CE / T-mark / PSE
Electrical
Topology Switching Flyback
Dielectric Withstand 4,000VAC Primary - Secondary
Touch Current (Patient) < 100µA
Leakage Current < 100µA
Earth Leakage Current < 5mA
Efficiency Meet DoE Level VI, ErP Stage 2
EMC Standards EN55011,
EN61000-3-2,3,
EN61000-4-2,3,4,5,6,8,11
MTBF 300,000 Calculated Hours at 25°C, by Telcordia SR-332
Environmental
Operating Temperature -20 to +40°C
Storage Temperature -20 to +80°C
Operating Altitude 5,000M
Relative Humidity Operating: 20 to 80% RH,
Storage: 10 to 90% RH
Cooling Natural Convection Cooling
Mechanical
Case Dimension C14 & C18 - L 100 x W 50 x H 33 (mm)
C8 - L 99 x W 50 x H 33 (mm)
Weight 210g (Ref.)

Overview

ATM036T‑P/A Series 36W Desktop Medical Power Adapter

The ATM036T‑P/A series is a 36W medical desktop AC/DC power supply designed for healthcare and clinical devices that demand reliable, regulated, and safe external power. Built to comply with IEC/EN/ANSI/AAMI ES60601-1 standards, this series delivers stable DC output for sensitive medical electronics while ensuring patient and equipment safety.

Its desktop design allows flexible placement on workstations, shelves, or inside medical cabinets, while multiple DC connector and cable options enable seamless integration into different medical devices. The series supports global deployment with universal AC input and configurable plug options.

Reliable Performance for Sensitive Equipment

Engineered for continuous operation, the ATM036T‑P/A delivers a low-noise, regulated output to maintain accurate device performance. Built-in protections—overcurrent, overvoltage, short-circuit safeguards, and 2 MOPP isolation—ensure both patient safety and device longevity.

Energy-efficient design reduces standby power consumption, making it ideal for devices that operate for extended periods, including patient monitors, portable diagnostic instruments, and laboratory equipment.

Applications and Use Cases

The ATM036T‑P/A series is suitable for:

  • Patient monitoring systems and bedside medical devices
  • Portable clinical instruments and diagnostic tools
  • Laboratory devices requiring regulated, reliable power
  • Equipment that benefits from a medical AC DC power supply with minimal electrical interference

These applications gain from its consistent output, low ripple, and enhanced isolation, ensuring accurate readings and reliable operation in clinical settings.

Support and Warranty

The series comes with a one-year warranty and technical support for device integration and custom project requirements. Options for DC connector type, cable length, and plug configuration are available to match specific device designs, ensuring compliance with medical standards.

FAQ

Q1: What devices can use the ATM036T‑P/A series?
A1: Suitable for patient monitors, portable diagnostic instruments, laboratory devices, and other medical electronics needing regulated, safe, and continuous power.

Q2: What safety standards does it meet?
A2: Fully compliant with IEC/EN/ANSI/AAMI ES60601-1, providing 2 MOPP isolation to safeguard both patients and connected devices.

Q3: Can this desktop power adapter be used internationally?
A3: Yes. Universal AC input (100–240 VAC) and configurable DC outputs allow deployment in global markets using appropriate AC power cords.

Q4: Can the DC connector or cable be customized?
A4: Yes, the series supports custom DC connectors, cable lengths, and plug configurations for project-specific requirements.