製品
300 Watt Medical Series - Desktop Type
商品番号: ATM300T-□XXX XXX = O/P Voltage
□ = P : C14□ = A : C18
Download Catalogue 機能:
* IEC/EN/ANSI/AAMI ES 60601-1 (ED 3.1) * EMC: IEC60601-1-2:2014 (ED 4.0) * IEC 60950 approval * Means of Protection: 2 X MOPP * Touch Current: < 100µA * 100-240VAC Universal Input * Single Output to 300W * Regulated Output with Low Ripple Noise * Safety Agency Requirements and EMI/EMS Certified * Private Label Marking Available * Modified and Custom Design Available * 2 Year Warranty
Green Mode
Meet CEC, DoE Level VI, Energy Star, ErP Stage 2
No Load Power Consumption Less Than 0.5W
Model
O/P Voltage
O/P Current
Watt
ATM300T-□120
12.0V
25.00A
300W
ATM300T-□150
15.0V
20.00A
300W
ATM300T-□240
24.0V
12.50A
300W
ATM300T-□480
48.0V
6.25A
300W
Input
Voltage
100-240 VAC
Line Frequency
50-60Hz
Current
3.5A - 1.5 A
Protection
Internal Primary Current Fuse
Configuration
IEC60320/C14, C18
Output
Load Regulation
± 5%
Ripple & Noise
1% Vp-p Max. for Output Voltage @Full Load
Transient Response
0.5ms for 50% Load Change Typical
Hold-up Time
10mS @ Full Load
Protection
Short Circuit Protection/ Over Voltage Protection/ Over Current Protection/ Over Temperature Protection
Ferrite Core
Yes
Safety Approvals
CB / UL / cUL / FCC / CE / T-mark / PSE
Electrical
Topology
LLC
Dielectric Withstand
4000 VAC Primary - Secondary
Touch Current
< 100µA
Efficiency
Meet DoE Level VI
EMC Standards
EN55011
EN61000-3-2,3
EN61000-4-2,3,4,5,6,8,11
MTBF
300,000 Calculated Hours at 25°C, by Telcordia SR-332
Environmental
Operaturing Temperature
0 to +40°C
Storage Temperature
-20 to +80°C
Relative Humidity
Operating: 20 to 80% RH
Storage: 10 to 90% RH
Cooling
Natural Convection Cooling
Mechanical
Case Dimension
L 254 x W 116 x H 47 (mm)
Weight
1600g (Ref.)
注文情報:
最低発注量: Negotiable
OEM/ODM Product, Branded Product, Buyer's Label Offered
FOB(本船渡し): HK